Olymps FluoView FV1200 - Multi-Area Time-Lapse Imaging
Olymps FluoView FV1200 - Multi-Area Time-Lapse Imaging
- For more advanced image acquisition configurations with multiple positions set up your channels as described in SIMPLE CONFOCAL IMAGE ACQUISITION. Then open the Multi-Area Time-Lapse controller, which can be found under Device -> Multiarea Time-Lapse.
- An XY-stage calibration warning will open. Press OK to calibrate the stage, which is important to be able to register points for multipoint acquisitions.
- If you would like to use the Z-Drift Controller (ZDC) to keep the focus stable throughout the experiment follow the instruction in the Z-DRIFT CORRECTION (ZDC) section. Be aware that you will need to restart the software to set this up.
- In Multiarea Time-Lapse Controller:
- 1st button from left: Add current position (to save a position, then move on to next -> gives you list of data points in Registered Point List)
- 2nd button from left: Mosaic Outline (to select separate points and then fills in gaps between them to scan an area)
- 3rd button from left: Define Matrix (specify central point and define area around it)
- Go to View -> Stage controller to control the stage using the software or use the manual XY-controller to navigate to different positions in your sample and create a point map using the Add Current Position, Mosaic Outline or Define Matrix method.
- Double-click on a number in the Registered Point List to go to positions you marked previously.
- In the Add Current Position mode it is possible that the added points have different Z-positions.
- The Interval and Repeat field allow you to set-up time series.
- It is also possible to combine the multi-point acquisition with a Z-stack.
- To reload a previous point list navigate to Device -> Multi-Area Time-Lapse Viewer -> Open files in the main window.
- To clear points go to the Registered Point List, highlight points you do not want to keep and press the trash button.
- When you want to acquire a Z-stack or time-series you need to go to Edit -> Edit acquisition parameters and select Depth and/or time respectively. Then press OK. Be aware that this only activates the Z-stack for the currently selected position and needs to be activated for all other positions separately.
- If you make changes to a marked position and want to update it go to Edit -> Re-register.
- Before starting the image acquisition create a location on the PC hard disk to save your data to by navigating to Options -> File & Folder -> Folder name (you can also create a subfolder for each data point).
- When you are happy with the experiment set-up press the Ready button (bottom left) in the Registered Point List window.
- Press Start to start the image acquisition.
- To obtain a stitched image from raw Mosaic Outline or Define Matrix data go to Stitch View -> select central plane if Z-stack -> Show -> Save as series.