Multiphoton Lasers and Fluorescence Filters
Lasers
This microscope is equipped with a 405nm diode, a 561nm DPSS, a 633 HeNe, and an Argon laser with lines at 458nm, 488nm and 514nm. It also has a Ti:Sapphire laser tuneable between the wavelengths of 720-1064nm, and the OPO is tuneable between 1050-1300nm.
Dichroic Mirrors for Invisible Light
| Position | Dichroic | Example Fluorophores | Notes |
|---|---|---|---|
| 1 | Plate | Use this when not using any Invisible Light lasers | |
| 2 | Mirror | ||
| 3 | MBS 405 | DAPI, Hoechst | |
| 4 | MBS 445 | Not used | |
| 5 | MBS 690 | Various | Use this for multiphoton laser ablation at <760 nm as well as imaging |
| 6 | MBS 760 | Various | Use this for multiphoton imaging >760 nm |
| 7 | MBS 405(UV)/760 | Various | Use this for combining excitation with the 405 nm laser with the multiphoton (>760 nm) |
| 8 | MBS 445/760 | Not used | |
| 9 | MBS T80/R20 | Use for reflection imaging | |
| 10 | None |
Dichroic Mirrors for Visible Light
| Position | Dichroic | Example Fluorophores | Notes |
|---|---|---|---|
| 1 | MBS 458 | CFP | |
| 2 | MBS 458/514 | CFP/YFP | |
| 3 | MBS 458/561 | CFP/RFP | |
| 4 | None | ||
| 5 | MBS 488 | GFP, Alexa 488 | |
| 6 | MBS 488/561 | GFP/RFP, Alexa Fluor 488/Alexa Fluor 568, FITC/TRITC | |
| 7 | None | ||
| 8 | MBS 488/561/633 | GRP/RFP/Alexa Fluor 633 | |
| 9 | MBS T80/R20 | Use for reflection imaging | |
| 10 | Plate |
Filter sets in the microscope stand reflector revolver
| Position | Filter Set | Notes |
|---|---|---|
| 1 | FSet49 wf | Zeiss Filter Set 49. Bandpass filter set for viewing DAPI and similar blue fluorophores down the eyepieces. |
| 2 | FSet38 wf | Zeiss Filter Set 38. Bandpass filter set for viewing Alexa 488, eGFP and similar green fluorophores down the eyepieces. |
| 3 | FSet43 wf | Zeiss Filter Set 43. Bandpass filter set for viewing TRITC, DsRed and similar orange fluorophores down the eyepieces. |
| 4 | BS-MP 355/690+(R) | Only for use with multiphoton. Reflects laser light >690 nm, and transmits light <690 nm to the non-descanned detectors. |
| 5 | BS-MP -760 | Only for use with multiphoton. Reflects laser light >760 nm, and transmits light <760 nm to the non-descanned detectors. |
| 6 | Rear | For confocal imaging. Sends all light to the scan head. |
, multiple selections available,