Hardware

  1. Turn the green switches (PC Microscope – Scanner Power – Laser Power) on the right side of the desk from left to right on and turn the key in the on position to switch on the microscope.



  2. Make sure that the white light laser (WLL) key is in ON position (always leave on).

     

  3. To turn on the STED laser press the button and wait until the orange LED (A) turns green (B, SHG ready). This should not take longer than approx. 10 min. When the LED turns green turn the key in the ON position.



  4. Turn on the FL6000 epi-fluorescence light source.



  5. Login to the PC using the TCS-User account. No password is required.


Software

 


  1. Double click on the LAS X desktop icon to start the Leica acquisition software.



  2. A Configuration pop-up window opens. Select Configuration - machine.xlhw, Microscope – DMI6000 and Resonant – OFF (unless you want to use the resonant scanner for fast imaging).
    1. For the confocal only mode select STED – OFF
    2. For the STED mode select STED – ON

    Press OK.



  3. When the Microscope Stand pop-up window appears make sure that the condenser is all the way up, so that it does not interfere with the moving stage, then select YES). 



  4. After some time the start screen of the software will appear and you are ready to start setting up your experiment.

...