Lasers

Excitation Lasers
WavelengthTypeFluorophores
405 nmCWDAPI, Hoechst, etc
White Light Laser†PulsedAll visible dyes from green to far red
STED Depletion Lasers‡
592 nmCWSTED with green dyes
660 nmCWSTED with red dyes
775 nmPulsedSTED with far red dyes

† The White Light Laser (WLL) is a super-continuum that can be tuned from 470 nm to 670 nm and can therefore be used to excite all visible dyes.

‡ The depletion lasers are used to increase resolution by stimulating re-emission of photons. They are not used for excitation and are only used in STED.


Objective lenses

Objective lenses
MagnificationTypeImmersionNAWorking Distance (mm)Coverslip correction
10xHC PL FLUOTARAir0.3110.17
20xHC PL APO CS2Imm (Oil, Glycerol and Water)0.750.680/0.17
25xHC FLUOTAR L VISIRWater0.952.40.17
40xHC PL APO CS2Oil1.30.240.17
63xHC PL APO CS2Oil1.40.140.17
100xHC PL APO CS2 (dedicated STED objective)Oil1.40.130.17

NOTE: There is only space for 6 objective lenses on the objective turret, so if the lens you would like to use is not on there please contact lmcb-lm-help@ucl.ac.uk




MICROSCOPE: DMI6000, inverted


 

STED RESOLUTION 

XY

~ 50 nm

Z

same as confocal



IMAGING MODES:

  • Laser scanning confocal imaging (multi-colour, multi-point, live cell)
  • STED with 592 nm depletion, up to 3 colours



FLUORESCENCE FILTER SETS: This microscope uses an Acousto-Optical Beam Splitter (AOBS) with a freely selectable emission detection range rather than fluorescence emission filter cubes.



DETECTORS


3x HyD, high sensitivity, 3 operating modes (standard, counting, BrightR) + time-gating

3x PMT, one of which for transmitted light imaging

A


INCUBATION: The microscope is fully enclosed and equipped with The Cube chamber heater and The Brick CO2 gas mixer.



SOFTWARE 

Acquistion

LAS X

Deconvolution

Huygens Professional Version 15.10.1p2

A


ADDITIONAL INFORMATION:

  • resonant scanner (8000 Hz) for fast imaging
  • anti-vibration optical table
  • fully enclosed
  • autofocus
  • motorized XY stage
  • Z-galvo