  1. When you have finished with your experiments first make sure that all your data have been saved in a secure place. Also make a backup of them.

  2. Remove your sample from the microscope stage and clean all objective lenses you have been using thoroughly with the Whatman lens cleaning tissue provided. Also make sure to check the nosepiece for oil spills and if there are any clean them up.

  3. Close the NIS-Elements software (If someone else is booked on the microscope after you you can stop here, otherwise proceed with step 4 - 6).

  4. Shut down the PC.

  5. On the front of the Agilent laser bed switch off the four laser line buttons and turn the key in the OFF position.

  6. Last switch off the two sockets on the wall labelled Laser Bed and N-STORM Devices.