  1. These instructions assume that you have already switched on all components of the microscope and placed your specimen on the microscope stage and focused on it using the ocular. If focusing through the ocular is not possible e.g. because your sample is labelled with a far red dye, you can also do this using the live preview mode in the NIS-Elements software.

  2. Remember to select Andor with N-STORM in the camera driver selection dialog when starting up the software, so the Andor iXon DU-897 EMCCD camera required for STORM acquisitions will be available in the software

  3. At the bottom left of the NIS-Elements software navigate to the N-STORM acquisition tab, which opens a software layout containing all controls required for the acquisition of STORM images.