The image above shows condensers of various upright and inverted microscopes. This selection is not exhaustive and components may be located elsewhere on other models.
- Field diaphragm adjustment (A)
...
- : The DM2500 field diaphragm is on the other side of the stand. The Nikon Ti field diaphragm is near the top of the transmitted light pillar, labelled F
...
- Condenser focus (B)
...
- : This moves the condenser up and down but doesn't affect the focus of the specimen
...
- Centring screws
...
- (C)
...
- : The Leica DM2500 and DMI6000 must be centred using hexagonal screwdrivers or Allen keys
...
- Iris aperture adjustment (D)
...
- : The iris aperture on the DMI6000 is adjusted using the AP buttons on the left of the microscope stand
Alignment Procedure
- Focus on a transmitted light image of the specimen using a 10x air lens
- Rotate the condenser turret to the bright field position. This will be labelled H on Zeiss microscopes, A on Nikon microscopes and usually BF on Leica microscopes. Adjustment can either be manual or done through buttons or a touchpad, depending on the microscope model
- Fully close the field diaphragm and adjust the condenser focus so the image of the iris is in sharp focus
- Centre the image of the iris using the two centring screws
- Open the field diaphragm until it is just outside the field of view
- Remove an eyepiece and look into the empty opening
- Adjust the iris aperture until it is about 70% open and then replace the eyepiece. The microscope is now adjusted for Köhler illumination
...