Click the Z tab to set the range and step size for Z series. Click on the ‘Defined by top bottom’ button (see red arrow below) to set a top and a bottom for the Z series. Set the Step size. Clicking the button next to the Step field will set the step to the optimum for the objective lens. For multi-position acquisitions the Z series must be Relative to compensate for the different Z positions of the marked points. Click the ‘Symmetric mode defined by range’ button next to 'Defined by top bottom' and click Relative to make a symmetric z series around every point whose range is defined by the top and bottom you set above. The third button for ‘Asymmetric mode defined by range’ is more difficult to set up since you need to be able to type in the range above and below the Relative home position. Use the Order of Experiment button to define the order of acquisition for channels and Z slices.

Z tab in ND Acquisition

Setting up the length of the timelapse and frequency of acquisition

Click the Timing… button to check how long a single loop will take; this will determine the minimum Interval between time points. Click the Time tab and select at least one tick box at the start of a row. Specify the desired Interval and a Duration or number of Loops. Ticking a box in the row below will create another time Phase, so for instance you can have a fast acquisition for a short period followed by a slow acquisition over a longer period. Click the ‘Close Active Shutter when Idle’ box to reduce photobleaching.

Time tab in ND AcquisitionImage Added

Click ‘1 time loop’ to test the settings or ‘Run Now’ to begin the timelapse.