  1. Click the View button in the Acquisition Control area to begin a live feed from the acquisition camera.
  2. There will be two live images, which correspond to two simultaneous colour channels. For dSTORM using a far-red dye you only need Channel 1, so you can untick Channel 0 in the Image Display Options on the left. For dyes emitting wavelengths lower than far-red you will need Channel 0 to be ticked (but possibly not Channel 1 if you are not using a far red dye).
  3. The live camera image will only show noise until you illuminate the sample with a laser. Change the percentage power in the Light Control area by either clicking on the spinner or typing in a value and pressing the Return key. The laser power in the image below is set to 100% for dSTORM acquisition, but it is best to use a lower power like 10% to avoid photobleaching while setting up your specimen. Click on the button above the laser (e.g. in this case the red 640 button) to switch on the illumination.
  4. Focus on your specimen. If you are using Z lock you may need to Z Lock is NOT activated then you should change the value in the Z: field to move the focus. If Z lock is active the Z: position cannot be changed, so you should adjust the Z offset as explained in the section on focusing. Both the Z field and Z offset are highlighted with a red outline in the Position Control image below.
    Image Added
  5. Once you're in focus turn off the laser so you don’t photobleach the specimen.

...