Super-Resolution Microscopy

Sample Preparation

Glox Switching Buffer

The Glox buffer formulation is adapted from the following publication:

...

  • 110 µl Solution A
  • 880 µl Solution B
  • 220 µl Solution C
  • 990 µl PBS

Solution A (Enzyme Solution)

  • 10 µl catalase
  • 20 µl 1M TCEP
  • 2 ml glycerol
  • 125 µl 1M KCl
  • 100 µl 1M Tris-Cl pH 7.5
  • 50 mg glucose oxidase
  • Top up to 5 ml with distilled water

Dispense into 50 µl aliquots and store frozen at -20°C.

Solution B (Glucose Solution)

  • 4 g glucose
  • 4 ml glycerol
  • 36 ml distilled water

Dispense into 400 µl aliquots and store at -20°C.

Solution C (Reducing Agent Solution)

  • 113.6 mg MEA-HCl
  • 1 ml distilled water

Use fresh or make 100 µl aliquots for storage at -20°C

Sealing a 4-well ibidi chamber for STORM

This video shows how to fill one well of a 4-well ibidi with Glox and then seal the top with a coverslip so there are no air bubbles (and hence no oxygen ingress). It is also possible to seal the chamber by inverting the lid of the ibidi chamber and sliding that over the wells. In that way all the wells can be sealed. In the image below I have sealed the middle two wells by sliding the inverted lid across them.

...