Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

A Bio-Rad Radiance confocal microscope mounted on a Nikon E800 manual upright microscope stand. You can use this microscope for imaging slides and for specimens that can only be imaged in open chambers in aqueous media (e.g. tissue, pupae/embryos, fish) or that require a longer working distance. The optional water dipping lenses have working distances in the order of 2 mm. Live mammalian cells can be imaged if the imaging chamber can be inverted so the cells are upside down. There is an optional heated stage adapter for warming the specimen but there is no CO2 supply so the chamber must be sealed to prevent pH change. There are three lasers with six laser lines but there is no 405 nm laser, so you cannot image DAPI or Hoechst. The confocal has three GaAsP detectors for high sensitivity imaging of poorly fluorescent specimens or at low laser power. FRAP and other photo-stimulation experiments are possible

Table of Contents
minLevel2

Access & Usage

Technical Specifications

Objective

...

Lenses

Standard Objective Lenses
MagnificationTypeImmersionNADICWD (mm)
10xPlan FluorAir0.3-16
20xPlan FluorMImm0.75M0.33 - 0.35
40xPlan FluorOil1.3H0.2
60xPlan ApoOil1.4H0.21
Water Dipping Lenses
MagnificationTypeImmersionNADICWD (mm)
20xFluorWater0.5M2
40xFluorWater0.8M2
60xFluorWater1.0H2

The letters in column 5 indicate the correct prism for use with the lens

...