Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The microscope has 20x and 40x long working distance objective lenses that have adjustable correction collars so the objectives can be corrected to focus through different thicknesses. For example, coverslips are usually only 170 microns thick but the growth surface of standard cell culture dishes is approximately 1 mm thick, so a correction collar would have to be set differently for specimens mounted on one or other of these surfaces.

Table of Contents
minLevel2

Objective lenses

Magnification

Immersion

Designation

NA

Contrast Technique

Coverslip Correction (mm)

WD (mm)

10x

Air

N Plan

0.25

Ph1

None

17.6

20x

Air

N Plan LWD

0.4

Ph1

0 - 2

3.2 - 1.9

40x

Air

N Plan LWD

0.55

Ph2

0 - 2

3.3 - 1.9

20x

Air

N Plan LWD

0.4

LMC

0 - 2

3.2 - 1.9

40x

Air

N Plan LWD

0.55

LMC

0 - 2

3.3 - 1.9

...