Axio Imager
...
Wiki Markup |
---|
h1. Axio Imager ---- {float:right}!Axio-Imager.jpg|border=1!{float} * |
Location:
...
MRC
...
Building,
...
First
...
Floor,
...
Room
...
1.19.
...
The
...
Zeiss
...
Axio
...
Imager
...
upright
...
microscope
...
is
...
ideal
...
for
...
viewing
...
slide-mounted
...
specimens
...
using
...
phase
...
contrast
...
or
...
fluorescence
...
microscopy.
...
It
...
has
...
a
...
QImaging
...
Retiga
...
greyscale
...
camera
...
suitable
...
for
...
low
...
light
...
level
...
imaging
...
of
...
fluorescent
...
specimens.
...
The
...
objective
...
lenses
...
are
...
all
...
designed
...
for
...
high
...
resolution
...
observation
...
of
...
specimens
...
mounted
...
under
...
standard
...
#1.5
...
coverslips
...
(
...
170 µm).
...
All
...
microscope
...
controls
...
are
...
manual
...
apart
...
from
...
camera
...
image
...
capture.
...
There
...
are
...
currently
...
three
...
image
...
acquisition
...
software
...
packages
...
installed
...
on
...
the
...
computer:
...
Openlab
...
(obsolete
...
-
...
updates
...
no
...
longer
...
available);
...
Volocity
...
and
...
µManager.
...
To
...
request
...
training,
...
report
...
problems
...
or
...
ask
...
for
...
help
...
please
...
...
...
.
Table of Contents | ||
---|---|---|
|
Objective lenses
Magnification | Immersion | Designation | Contrast Technique | Coverslip Correction (mm) | WD (mm) | |
---|---|---|---|---|---|---|
10X | Air | EC Plan Neofluar | 0.3 | Ph1 | 0.17 | 5.5 |
20X | Air | EC Plan Neofluar | 0.5 | Ph2 | 0.17 | 2.0 |
40X | Oil | EC Plan Neofluar | 1.3 | Ph3 | 0.17 | 0.21 |
63X | Oil | Plan Apochromat | 1.4 | Ph3 | 0.17 | 0.19 |
100X | Oil | EC Plan Neofluar | 1.3 | Ph3 | 0.17 | 0.2 |
Fluorescence filter sets
Filter Position | Filter Set | Excitation Filter | Dichroic Mirror | Emission Filter | Fluorophores |
---|---|---|---|---|---|
1 | Chroma ET-DAPI | 350/50 | 400LP | 460/50 | DAPI/Hoechst |
2 | DIC | N/A | N/A | Analyser | DIC |
3 | Empty | N/A | N/A | N/A | Empty |
4 | Chroma ET/GFP (FITC/Cy2) | 470/40 | 495LP | 525/50 | FITC/GFP |
5 | Chroma ET/mCherry/Texas Red | 560/40 | 585LP | 630/75 | Texas Red |
6 | Chroma Custom ET Cy5 Longpass | 620/60 | 660LP | 665LP | Cy5/Alexa 647 |
Note that the ET filter sets are extremely efficient and transmit more light than older filter sets. This goes for the excitation filters as well as the emission filters, so it is possible for specimens to photobleach rather quickly. Please use the intensity dial on the Prior Lumen light source to reduce the excitation intensity if bleaching is a problem.