What does binning mean? (Andor Technologies) 

Gain

Gain is the amplification factor applied to an input signal to adjust the amplitude of the signal output. In confocal imaging using PMTs, an input voltage signal may be multiplied by a positive gain factor to increase the voltage of the signal output and hence the brightness of the image on the monitor.  In CCD and CMOS cameras 1X gain is defined as the standard number of photoelectrons assigned per ADU (analogue to digital unit); 2X gain is half the number of photoelectrons assigned per ADU, etc. Since decreasing the gain results in fewer electrons per ADU, the camera is more sensitive to low light levels. 

Anchor
LWD
LWD

Long Working Distance (LWD) objective lenses

...

Many LWD objectives have correction collars that can be adjusted for a variety of specimen thicknesses. If a collar has been adjusted for a 1 mm thickness (e.g. the bottom of a plastic dish) it will not focus correctly through a glass coverslip and you will need to adjust the collar to a setting appropriate to a 170 micron thickness.

There is no systematic way of adjusting correction collars. The best way is to set the collar to approximately the correct position before observing your sample, then to focus on the sample, and finally to make fine adjustments to the collar position if necessary. The adjacent picture shows the correction collar of a Leica objective lens when adjusted for a 1 mm sample thickness. The black dot marks the position of the collar and the numbers around the objective barrel indicate the thickness for which the lens will be adjusted. The knurled ring above the dot should be used to rotate the collar. Unfortunately the Leica objectives don't have a mark for the position for a 170 micron coverslip, so the correct position can only be set approximately.

...

Numerical Aperture (Wikipedia)

Anchor
WD
WD
  

Point-spread function (PSF)

 

In fluorescence microscopy a point source of excitation light is focused on the specimen as a diffraction limited spot, which, in the case of a high NA oil objective (NA=1.4 or above) will have a diameter of 200-300 nm, depending on the excitation wavelength.  The blurred extended diffraction image is described by a point-spread function, and is a measure of image degradation.  It can be determined by taking image stacks through sub-resolution fluorescent beads mounted in the same medium and emitting at a similar wavelength as fluorophores as the specimen.  Alternatively, it can be calculated theoretically using the microscope control software or commercial algorithms.      

 

Working Distance (WD)

The working distance is the distance from the front of the lens to the nearest surface of the coverslip when the specimen is in sharp focus. For objective lenses designed for use without a coverslip the surface of the culture vessel or specimen in used.

...