Olympus FluoView FV1200 - Technical Specifications
Olympus FluoView FV1200 - Technical Specifications
MICROSCOPE: IX83, inverted
IMAGING MODES
- Laser scanning confocal imaging (multi-colour, multi-point, live cell time-lapse, Z-stack)
- FRET
- FRAP
- FLIP
- λ-scan
- uncaging
LIGHT SOURCES
Light source | Type | Wavelengths [nm] |
Diode | Laser | 405, 440 |
Diode | Laser | 559 |
Diode | Laser | 635 |
Multi-line Argon ion | Laser | 458, 488. 515 |
Ablation laser | Laser (pulsed or cw) | 405 |
U-HGLGPS | Mercury metal halide bulb | ~ 300 - 700 (peaks at 365, 405, 436, 546, 579) |
A
OBJECTIVE LENSES
Magnification | Immersion | Type | NA | Coverslip correction [mm] | WD [mm] |
---|---|---|---|---|---|
10x | air | UPLSAPO10X2 | 0.4 | 0.17 | 3.1 |
20x | air | UPLSAPO20X | 0.75 | 0.17 | 0.6 |
40x | oil | UPLFN40X0 | 1.3 | 0.17 | 0.2 |
60x | oil | PLAPON60X0SC2 | 1.4 | 0.17 | 0.12 |
60x | silicone oil | UPLSAPO60XS | 1.3 | 0.15 - 0.19 | 0.3 |
30x | silicone oil | UPLSAPO30XS | 1.05 | 0.13 - 0.19 | 0.8 |
60x* | water | UPLSAPO60XW | 1.2 | 0.13 - 0.21 | 0.28 |
NA = Numerical Aperture, WD = Working Distance |
There is also a 60x UPLSAPO60XW water lens with NA 1.2, WD 0.28 and a correction collar for coverslip thicknesses 0.13 to 0.21 mm. It is kept in a drawer because there isn't a free position in the nose piece. If you need to use this leans then please ask a member of the Light Microscopy Facility staff because the microscope software settings must be edited when an objective lens changes.
FLUORESCENCE FILTER SETS: Filter sets for imaging a wide range of standard visible and far-red fluorophores are available.
DETECTORS
- 3 channel PMT (Hamamatsu R7862)
- External transmitted light PMT (Hamamatsu R7400)
- Ultra-high sensitivity detector GaAsP with onboard Peltier cooling system (2 channel)
INCUBATION: TokaiHit CO2 stage top incubator including lens heater
SOFTWARE: FV10-ASW 4.2
ADDITIONAL INFORMATION
- anti-vibration optical table
- motorized Z-drift compensator (ZDC)
- motorized XY stage
- Silver coated galvo scanner
- SIM scanner for simultaneous photostimulation and confocal image acquisition