Remove the dust sheet from the microscope stand and put it in the box beneath the table. Turn on the rocker switch on the power distribution unit behind the microscope stand. This supplies power to the PC and monitor as well as the Examiner and WSB PiezoDrive control boxes on the shelf above the microscope stand. Turn on the System and Components buttons on the Component Rack. Turn the Laser key one quarter-turn clockwise from 0 to I. Wait for the Laser ON LED to go off; there will hear a short sequence of beeps when this happens. Turn on the PC. Note that you must wait for the LED to go off before you boot the PC, otherwise the software might not recognise the components.