  1. Remove the dust sheet from the microscope stand and put it in the box beneath the table.
  2. Turn on the main switch on the power distribution unit behind the microscope stand. This supplies power to the PC and monitor as well as the Examiner and WSB PiezoDrive control boxes on the shelf above the microscope stand.
    Image RemovedImage Added
  3. Turn on the System and Components buttons on the PSUbox in the Component Rack. Turn the Laser key in the LMboxone quarter-turn clockwise from 0 to I. The white Laser ON LED will come on for a short time after the key has been turned.
    Image RemovedImage Added
  4. Wait for the Laser ON LED to go off; there will hear a short sequence of beeps when this happens.

    Note

    You must wait for the Laser ON LED to go off before you boot the PC, otherwise the software might not recognise the components.


    Image RemovedImage Added

  5. Turn on the PC. Log into Windows. Double-click the Zen icon to open the software.

...