Note that the 6 mm working distance of the Clarity lens will mean that any sample imaged using this lens will most likely have to be mounted in a deep vessel.

Lasers

The DIVE has the followingsix visible solid state lasers for confocal microscopy.

Wavelength

Colour

Example dyes

405VioletDAPI, Hoechst
448IndigoCFP
488BlueGFP, Alexa 488
514TurquoiseYFP
561YellowmCherry, Alexa 568
638RedCy5, Alexa 647

The infrared laser for multiphoton microscopy is a Coherent Discovery NX with a fixed line at 1040 nm that can be used for red dyes such as tdTomato and mCherry, and a tuneable line that can be adjusted between 660 nm and 1320 nm to excite a range of dyes. The fixed and tuneable lines can be used simultaneously.