Oil | Refractive Index | For use with... | Kept in... |
---|---|---|---|
Cargille Type HF | 1.518 (at 23ºC) | Oil lenses at room temperature | Tray outside environmental chamber |
Cargille Type 37 | 1.518 (at 37°C) | Oil lenses at 37ºC | Tray inside environmental chamber |
Zeiss Immersol W | 1.334 (at 23ºC) | Water immersion lens only | Tray inside environmental chamber |
Note that Cargille Type 37 is optimal for use at 37 degrees but introduces spherical aberrations when imaging beyond anout 10 microns into aqueous media because it is not refractive indexed matched to water. Zeiss Immersol W in combination with the 60x WI lens might be a better choice in this case, but its RI is corrected for use at room temperature and won't be correct at 37 degrees.
Setting up the Nikon Perfect Focus System (PFS) for multi-point acquisition in Z
...