Description


Radiance confocal microscope

This is a Bio-Rad Radiance confocal microscope mounted on a Nikon E800 manual upright microscope stand. You can use this microscope for imaging slides and for specimens that can only be imaged in open chambers in aqueous media (e.g. tissue, pupae/embryos, fish). The optional water dipping lenses have working distances in the order of 2 mm. Live mammalian cells can be imaged if the imaging chamber can be inverted so the cells are upside down. There is an optional heated stage adapter for warming the specimen but there is no CO2 supply so the chamber must be sealed to prevent pH change. There are three lasers with six laser lines but there is no 405 nm laser, so you cannot image DAPI or Hoechst. The confocal has three GaAsP detectors for high sensitivity imaging of poorly fluorescent specimens or at low laser power. FRAP and other photo-stimulation experiments are possible

Training & Booking

Specifications

Objective Lenses

Standard Objective Lenses
MagnificationTypeImmersionNADICWD (mm)
10xPlan FluorAir0.3-16
20xPlan FluorMImm0.75M0.33 - 0.35
40xPlan FluorOil1.3H0.2
60xPlan ApoOil1.4H0.21
Water Dipping Lenses
MagnificationTypeImmersionNADICWD (mm)
20xFluorWater0.5M2
40xFluorWater0.8M2
60xFluorWater1.0H2

The letters in column 5 indicate the correct prism for use with the lens

Lasers

LineLaserTypical FluorophoresNotes
458ArgonCFPThis is not a very powerful laser line
476Argon-Could be used as an alternative for green dyes when trying to separate dyes with similar spectra
488ArgonEGFP, FITC, Alexa 488 
514ArgonYFPCould also be used for Citrine, Venus, etc
543GreenHeNeTRITC, Alexa 532, Alexa 546, Alexa 555, Cy3Can also be used for 'redder' dyes like Texas Red, Alexa 568, Alexa 594, mCherry but with lower excitation efficiency
637Red DiodeCy5, Alexa 633, Alexa 647, Si Rhodamine dyes, DRAQ5 

Note that there is no 405 nm laser on this machine so you cannot image Hoechst, DAPI or anything else in the blue region of the visible spectrum