MICROSCOPE: DMI6000, inverted


 

STED RESOLUTION 

XY

~ 50 nm

Z

same as confocal



IMAGING MODES:

  • Laser scanning confocal imaging (multi-colour, multi-point, live cell)
  • STED with 592 nm depletion, up to 3 colours


LIGHT SOURCES  

Light source

Type

Wavelengths [nm]

Diode laser

Laser

405

White light laser (WLL)

Laser

470 - 670, tunable

Depletion laser

Laser

592

Leica EL6000 

Mercury metal halide bulb

~ 300 - 700 (peaks at 365, 405, 436, 546, 579)

A


 

OBJECTIVE LENSES     

Objective

Magnification

Immersion Medium

Numerical Aperture

Coverslip correction [mm]

Working Distance [mm]

HCX PL APO CS10xair0.40.172.2
HC PL FLUOTAR10xair0.3none11
HC PL APO CS220xIMM (water, glycerol, oil)0.75with or without coverglass/correction collar0.68
HCX APO LU-V-I20xwater0.5N/A35
HC FLUOTAR L VISIR25xwater0.950.1724
HC PL APO CS240xoil1.30.170.24
HC PL APO CS2 /1.4 (WD = 0.14 mm)63xoil1.40.170.14
HC PL APO CS2 (dedicated STED objective)100xoil1.40.170.13

NOTE: There is only space for 6 objective lenses on the objective turret, so if the lens you would like to use is not on there please contact lmcb-lm-help@ucl.ac.uk



FLUORESCENCE FILTER SETS: This microscope uses an Acousto-Optical Beam Splitter (AOBS) with a freely selectable emission detection range rather than fluorescence emission filter cubes.



DETECTORS


3x HyD, high sensitivity, 3 operating modes (standard, counting, BrightR) + time-gating

3x PMT, one of which for transmitted light imaging

A


INCUBATION: The microscope is fully enclosed and equipped with The Cube chamber heater and The Brick CO2 gas mixer.



SOFTWARE 

Acquistion

LAS X

Deconvolution

Huygens Professional Version 15.10.1p2

A


ADDITIONAL INFORMATION:

  • resonant scanner (8000 Hz) for fast imaging
  • anti-vibration optical table
  • fully enclosed
  • autofocus
  • motorized XY stage
  • Z-galvo