Objective lenses
Magnification | Immersion | Type | DIC | NA | Cover slip correction | WD (mm) |
---|---|---|---|---|---|---|
10x | Air | Plan Apo λ | DIC L | 0.45 | 0.17 | 4 |
20x | Air | Plan Apo | DIC M | 0.75 | 0.17 | 1 |
40x | Air | Plan Apo | DIC M | 0.95 | 0.11 - 0.23 | 0.21 |
60xA | Oil | Plan Apo | DIC N2 | 1.4 | 0.17 | 0.21 |
100x | Oil | Plan Apo λ | - | 1.45 | 0.17 | 0.13 |
100x† | Silicone Oil | SR HP Plan Apo λS | - | 1.35 | 0.15 - 0.19 | 0.3 |
†The 100x silicone oil lens is available on request. NA = Numerical Aperture, WD = Working Distance |
Fluorescence filter sets
Wide-field Fluorescence
Sedat quad-band dichroic and single-band emission filters in the filter wheel on the right-hand port. Lower speed but reduced cross-talk between channels.
Filter ID | Excitation Wavelength | Dichroic Mirror | Emission Filter (in wheel) | Example Fluorophores |
---|---|---|---|---|
DAPI | 385 nm | 89000 Sedat Quad | 435 - 485 | DAPI, Hoechst |
GFP | 470 nm | 89000 Sedat Quad | 500 - 550 | GFP, FITC |
RFP | 550 nm | 89000 Sedat Quad | 590 - 650 | RFP, mCherry |
Cy5 | 660 nm | 89000 Sedat Quad | 662 - 737 | Cy5 |
CFP | 470 nm | 89002 CFP/YFP | 460 - 500 | CFP |
eYFP | 500 nm | 89002 CFP/YFP | 520 - 550 | YFP |
Quad-band filters for fast switching to the Zyla Camera. High speed but risk of cross-talk.
Filter ID | Excitation Wavelength | Dichroic Mirror | Emission Filter (in quad) | Example Fluorophores |
---|---|---|---|---|
DAPI | 385 nm | 86012v2 Dapi/FITC/TxRed/Cy5 | ~ 455nm - 485nm | DAPI, Hoechst |
FITC | 470 nm | 86012v2 Dapi/FITC/TxRed/Cy5 | ~ 510nm - 545nm | GFP, FITC |
TxRed | 550 nm | 86012v2 Dapi/FITC/TxRed/Cy5 | ~ 590nm - 630nm | RFP, mCherry |
Cy5 | 635 nm | 86012v2 Dapi/FITC/TxRed/Cy5 | ~ 670nm - 740nm | Cy5 |
iSIM
The iSIM has the following quad-band emission filter in its scan head: