The SPE3 is a semi-automated, inverted Leica TCS SPE confocal microscope with an automated stage and z galvanometer insert. It will take slides, dishes and plates, but please not that all culture-style vessels must have glass bottoms of standard #1.5 coverslip thickness (170 µm). The confocal scanning system has the following lasers:

Lasers and Detectors

  • 405 nm (violet, for blue dyes like DAPI)
  • 488 nm (blue, for green dyes like GFP)
  • 561 nm (lime green, for red dyes like mCherry)
  • 635 nm (red, for far red dyes like Cy5)

It has a quad dichroic mirror meaning a rapid switch between channels.

There is only one photomultiplier tube (PMT) so all multi-channel acquisition is done sequentially

...