SPE3 microscope

...

components

...

The SPE confocal system is attached to a DMI4000 microscope stand. The fluorescence filter turret and objective lens nose-piece are manually rotated but the positions are encoded so the software knows which objective lens is in place. The microscope stand focus in manual so confocal sectioning is achieved via a galvanometer driven stage insert (Z Galvo). Eyepiece or confocal light path selection is automated.

Table of Contents
minLevelmaxLevel3
maxLevelminLevel3

Objective lenses

...

Leica ACS objective lenses have been designed specifically for SPE confocal microscopes with 405 nm lasers to correct for colour registration errors between the violet and visible lines. They do not have a flat field so Leica recommends that the confocal zoom is set to ≥1.6.

...

Turret Position

Filter Set

Excitation Filter

Dichroic Mirror

Emission Filter

Example Fluorophores

1

A

BP340-380

RKP400

LP425

Hoechst/DAPI

2

I3

BP450-490

RKP510

LP515

FITC/GFP

3

N2.1

BP515-560

RKP580

LP590

Texas Red

4

BF

N/A

N/A

N/A

Transmitted Light

5

N/A

N/A

N/A

N/A

Empty

6

Pol

N/A

N/A

N/A

Polariser for DIC and polarised light microscopy

...