Introduction

Table of Contents
typeflat

Note
titlePlease read!

If you select one of the specialist lenses in Agendo, or require incubation or CO2, then you must book the microscope in working hours (0900 - 1700) and leave a 30-minute gap before and after your session so facility staff can be available to swap the necessary equipment. Facility staff will receive an email when you book the machine and will be alerted to your requirement.

Expand
titleTechnical information and user guides

Technical specifications (objective lenses, lasers, etc.)

Start-up procedure

Controlling the microscope with the STP8000

Mark & Find using LAS X Navigator

Creating a Focus Map in LAS X Navigator

Image Added

Location: Medawar Building. Room 2.19B.

The Leica Microsystems Stellaris 8 DIVE (Deep In-Vivo Explorer) has the following features:

  • A fixed-stage upright DM6 microscope stand
  • The choice of standard objective lenses and special long-working distance dipping lenses for imaging deep into live and cleared specimens
  • A dual-beam Coherent Discovery multiphoton laser with a tunable laser line and a fixed 1040 nm laser line
  • A range of visible lasers for confocal imaging
  • Spectrally tuneable imaging with non-descanned detectors using the 4Tune system.

Technical Specifications

Standard objective lenses

The objective lenses in the table below are installed in non-adjacent positions in the standard 6-position nosepiece for objective lenses with M25 threads. Objective lenses are not installed in all positions to prevent damage to adjacent lenses as the objective lens is focused.

...

There is also a 10x HC PL APO CS2 0.4 NA (P/N 506424) air lens that can be installed in the standard nosepiece if required.

Specialised objective lenses

The objective lenses below have a larger M32 thread. To install these lenses the standard nosepiece must be removed from the microscope and replaced with a single-position nosepiece. Hence only one of these lenses can be installed at any one time. They have motorised correction collars which can be adjusted from within the software to correct for refractive index mismatches.

...

Note that the 6 mm working distance of the Clarity lens will mean that any sample imaged using this lens will most likely have to be mounted in a deep vessel.

Lasers

The DIVE has the following visible solid state lasers for confocal microscopy.

...

  • .