SP5 Shut-down Procedure
SP5 Shut-down Procedure
Shut Down Procedure for TCS SP5
- Exit the software by selecting 'Exit' from the File menu.
- Shut down Windows by selecting the shut-down option from the 'Start' menu in the bottom-left corner.
- Once Windows has shut down, switch off the 'PC Microscope' and 'Scanner Power' buttons on the main switchboard.
- Turn the key from On to Off but leave the 'Laser Power' button on. A fan cools the Argon laser for ten minutes after shut-down. Switching the Laser POwer button off switches off the fan.
- Switch off the EL6000 fluorescent light source. This can be done at any point as the light source is not electronically connected to the system.
- Wipe any immersion medium off the objectives you've used.
- Cover the microscope with its dust sheet.
- After ten minutes, switch off the Laser Power button on the main switchboard